Skip to content
ValiTracAI

Calibration by Parameter

Pressure, mass, torque, volume, electrical and dimensional calibration.

Calibration by Parameter

Pressure gauge calibration: procedure, points and uncertainty

How a pressure gauge is actually calibrated — rising and falling series, hysteresis, the head correction people forget, and what the certificate has to tell you.

2026-09-13 · 5 min read
Calibration by Parameter

Balance and scale calibration: eccentricity, repeatability and minimum weight

What a proper weighing-instrument calibration covers beyond a few test weights — off-centre loading, repeatability at low load, and the minimum weight that decides what you may weigh at all.

2026-09-13 · 4 min read
Calibration by Parameter

Torque wrench calibration under ISO 6789:2017

The 2017 revision split the standard into two parts and changed what a conformity certificate may claim. What Part 1 and Part 2 each cover, and why a Part 1 certificate is not an uncertainty statement.

2026-09-13 · 3 min read
Calibration by Parameter

Pipette calibration by the gravimetric method, under ISO 8655

Weighing the water a pipette delivers is simple arithmetic wrapped around several corrections people skip — the Z factor, evaporation, and the difference between systematic and random error.

2026-09-13 · 3 min read
Calibration by Parameter

Multimeter calibration: what the specification actually promises

A DMM specification is a formula, not a number, and it changes with range, function and how long since the last adjustment. How to read one and what a calibration has to cover.

2026-09-13 · 3 min read
Calibration by Parameter

Dimensional calibration: calipers, micrometers and gauge blocks

Why 20 °C is written into dimensional standards, what thermal expansion does to a steel measurement, and how calipers and micrometers are actually checked.

2026-09-13 · 3 min read
Calibration by Parameter

pH meter calibration: buffers, slope and why temperature is not optional

pH is a measurement whose reference values are themselves assigned with uncertainty. What the slope and offset actually tell you about an electrode, and why a two-point calibration is the minimum.

2026-09-13 · 3 min read
Calibration by Parameter

Conductivity meter calibration and the cell constant

Conductivity calibration determines a geometric property of the cell, not an offset. What the cell constant is, why temperature compensation dominates the result, and what goes wrong at low conductivity.

2026-09-13 · 3 min read
Calibration by Parameter

Force calibration and tensile testing machine verification

ISO 376 calibrates the force proving instrument; ISO 7500-1 uses it to verify the machine. Why the two are separate, and what alignment does to a force measurement.

2026-09-13 · 3 min read
Calibration by Parameter

Hardness testing machine calibration: direct and indirect verification

Hardness is defined by its test method, so a hardness machine is verified twice — once on its components and once end-to-end on certified reference blocks.

2026-09-13 · 3 min read
Calibration by Parameter

Particle counter calibration for cleanroom classification

ISO 21501-4 sets what a light-scattering particle counter has to demonstrate. Why counting efficiency at the smallest channel is the specification that matters, and how flow rate error becomes concentration error.

2026-09-13 · 3 min read
Calibration by Parameter

Flow meter calibration: what the reading depends on besides flow

Flow meters are sensitive to the installation as much as to the fluid. Upstream straight lengths, the difference between volumetric and mass flow, and why a bench calibration can mislead.

2026-09-13 · 3 min read