Skip to content
ValiTracAI

Practical references for measurement teams.

Everything here is free, written against the cited standards, and usable today — no sign-up, no gated PDFs.

Pt100Try it
-50 °C0 °C100 °C200 °C
  • Class AA
  • Class A
  • Class B
100 °C
R 138.505 ΩdR/dt 0.3793 Ω/°CClass A ±0.35 °C
IEC 60751: R(t) from the Callendar-Van Dusen equation; bands are the class B, A and AA tolerances over -50 to 250 °C.

New guides, straight to your inbox

Occasional email when we publish a new guide, calculator or template. No newsletter filler, no sharing your address.

Subscribe and get $40 off the ISO/IEC 17025 documentation pack.

We never sell your data