Equipment requirements in ISO/IEC 17025 (clause 6.4)
Equipment must be fit for purpose, calibrated where it affects results, checked, protected from adjustment, and fully recorded — including software.
Clause 6.4 is the equipment clause. It covers measuring instruments, software, reference standards, reference materials and consumables. Equipment must be verified to meet requirements before being placed into service, and a calibration programme must be established and reviewed to maintain confidence in calibration status.
Records the clause asks for
- Identity of the item and its software/firmware version.
- Manufacturer, model, serial number or other unique identifier.
- Evidence of verification that it meets requirements.
- Current location.
- Calibration dates, results, acceptance criteria and the date the next calibration is due (or the interval).
- Reference-material documentation.
- Maintenance plan and maintenance performed.
- Damage, malfunction, modification or repair.
When equipment is found out of tolerance, the laboratory must examine the effect on previous results and apply nonconforming-work procedures where needed.
Fit for purpose is a decision, and it has to be recorded
Clause 6.4.4 requires equipment to be verified as conforming to specified requirements before being placed or returned into service. That is not the same as unpacking it and checking it powers on. The specified requirements come from the method and the uncertainty budget: if the budget assumes a readout contributing 0.003 °C, verification means establishing that this readout does, not that the brochure says so.
The same applies on return from repair or recalibration. An instrument comes back adjusted, and the budget built on its previous behaviour no longer describes it. Reading the certificate before the instrument goes back on the bench is the control, and skipping it is how an out-of-tolerance finding stays undetected for a full interval.
Intermediate checks, which the clause requires and most laboratories skip
Clause 6.4.10 says that when intermediate checks are necessary to maintain confidence in performance, they must be carried out according to a procedure. For a reference thermometer calibrated annually, the honest question is what happens if it drifts in month three: without a check, every certificate for nine months is at risk, and a discovery at recalibration triggers clause 7.10 across all of them.
| Equipment | Check | Frequency |
|---|---|---|
| Reference PRT | Ice point, or a check standard measured and trended | Monthly, or before critical work |
| Readout or bridge | A stable reference resistor of known value | Monthly |
| Dry block | Compare against the reference at one set point | Quarterly |
| Working thermometers | Compare against the reference at one point | Between calibrations, at an interval set by their history |
Trend the results rather than filing them. A single ice-point reading proves very little; twelve of them plotted show drift long before the annual certificate does, and that chart is also the evidence for setting the interval under 6.4.7.
Software and firmware count as equipment
- The version is a required record under 6.4.13. A result cannot be tied to the calculation that produced it otherwise.
- A firmware update is a change to a verified item, and requires verification that it still meets requirements — including that stored coefficients survived the update.
- Equipment must be safeguarded from adjustments that would invalidate results (6.4.7): password-protected calibration menus, seals, or controlled access to coefficient entry.
- Spreadsheets used for calculation are in scope too, with version control and formula verification.
When equipment is found out of tolerance, clause 6.4.9 requires it to be taken out of service, labelled, and the effect on previous results examined. The look-back is the obligation people forget, and it is the one that costs the most when it is done late.
Frequently asked questions
- Must every thermometer in the laboratory be calibrated?
- Only those whose accuracy or uncertainty affects the validity of results or that are needed for traceability. Others need to be fit for purpose, which may be shown by a simpler check.
References
- [1]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories
General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.
Have a question on this topic?
Ask ValiTrac AI and see the evidence and calculation behind the answer.
Ask ValiTrac AIFree calculator: Certificate checker
Paste the text of a calibration certificate and screen it against ISO/IEC 17025:2017 clause 7.8: uncertainty with k, traceability, as-found and as-left, the decision rule behind a pass, and every other required element, with the matched text shown for each.
Open the calculatorCalibration on ValiTrac
Support calibration workflows, uncertainty budgets and certificate generation.
Open the calibration worksheetRelated articles
How to set calibration intervals
There is no universal interval. Set it from risk, manufacturer data and — above all — the instrument's own as-found history.
Metrological traceability in ISO/IEC 17025 (clause 6.5)
Results must be traceable to the SI through a competent laboratory's calibration, the certified value of a reference material, or direct realisation.