Skip to content
VTValiTracAI
ISO/IEC 17025

Equipment requirements in ISO/IEC 17025 (clause 6.4)

Equipment must be fit for purpose, calibrated where it affects results, checked, protected from adjustment, and fully recorded — including software.

ValiTrac AI editorialUpdated 2026-09-132 min read

Clause 6.4 is the equipment clause. It covers measuring instruments, software, reference standards, reference materials and consumables. Equipment must be verified to meet requirements before being placed into service, and a calibration programme must be established and reviewed to maintain confidence in calibration status.

Records the clause asks for

  • Identity of the item and its software/firmware version.
  • Manufacturer, model, serial number or other unique identifier.
  • Evidence of verification that it meets requirements.
  • Current location.
  • Calibration dates, results, acceptance criteria and the date the next calibration is due (or the interval).
  • Reference-material documentation.
  • Maintenance plan and maintenance performed.
  • Damage, malfunction, modification or repair.

When equipment is found out of tolerance, the laboratory must examine the effect on previous results and apply nonconforming-work procedures where needed.

Frequently asked questions

Must every thermometer in the laboratory be calibrated?
Only those whose accuracy or uncertainty affects the validity of results or that are needed for traceability. Others need to be fit for purpose, which may be shown by a simpler check.

References

  1. [1]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories

General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.

Related articles