Metrological traceability in ISO/IEC 17025 (clause 6.5)

Results must be traceable to the SI through a competent laboratory's calibration, the certified value of a reference material, or direct realisation.

ISO/IEC 17025ValiTrac AI editorialUpdated 2026-10-093 min read

Clause 6.5 requires the laboratory to establish and maintain metrological traceability of its measurement results by means of a documented unbroken chain of calibrations, each contributing to the measurement uncertainty, linking them to an appropriate reference.

Accepted routes to the SI

  • Calibration provided by a competent laboratory — in practice, one accredited under the ILAC arrangement or an NMI.
  • Certified values of certified reference materials from a competent producer.
  • Direct realisation of SI units ensured by comparison with national or international standards.

ILAC-P10 sets the accreditation-body policy that gives these routes teeth: a certificate from a non-accredited laboratory generally cannot be used to demonstrate traceability without the laboratory itself demonstrating that provider's competence.

Reading a certificate for traceability

Traceability is a property of a chain, and the chain is only as good as its weakest documented link. Four things have to be present on a certificate before it can support a traceability claim, and a certificate missing any of them is a gap you will have to close yourself.

What has to be on the certificate, and what it proves
ElementWhy it matters
An accreditation symbol or a named accreditation body and numberDistinguishes an accredited calibration from a non-accredited one done by the same laboratory on the same equipment
The measurement uncertainty, with its coverage factorEvery link in the chain must contribute to the uncertainty. A link with no stated uncertainty breaks the chain
Identification of the reference standards usedContinues the chain upward; a bare statement of traceable to national standards names no link at all
The measurand, conditions and points calibratedA chain is only unbroken for the quantity and range that was actually measured
What has to be on the certificate, and what it proves

The scope is narrower than the laboratory

A laboratory is not accredited; a scope is. The same laboratory may hold accreditation for Pt100 calibration from −40 °C to 250 °C and none at all for humidity, and it will happily perform the humidity work on a plain certificate. Check the point, the range and the quantity against the published scope rather than against the letterhead. This is also where an expired accreditation hides: the symbol is printed on the template long after the body has suspended it.

When you use a non-accredited supplier

ILAC-P10 is the policy that gives clause 6.5 teeth, and it does not absolutely forbid a non-accredited provider — it moves the burden onto you. If a laboratory uses a calibration from an unaccredited source to support traceability, it must itself demonstrate that provider's competence, traceability and uncertainty. In practice that means an assessment of the supplier, a review of their method and budget, and evidence of their own traceable references. That is a serious piece of work, and it is almost always cheaper to use an accredited supplier.

Where the chain usually breaks

  • An in-house reference used to calibrate working instruments, but never itself sent out. The chain stops at your own bench.
  • Certificates filed but never read, so nobody noticed that the calibration covered 0–100 °C and the instrument is used at −30 °C.
  • A statement of traceability with no uncertainty, which cannot contribute to a budget and therefore cannot form a link.
  • Reference standards recalibrated late, leaving a period during which every result they supported has an undocumented gap.
  • Adjustment at recalibration without as-found data, which severs the link to everything measured since the last certificate.

Frequently asked questions

Is a manufacturer's factory calibration acceptable for traceability?
Only if the manufacturer's laboratory is accredited for that calibration, or if you can demonstrate its competence yourself — which is rarely practical.

References

  1. [1]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories
  2. [2]ILAC-P10: ILAC Policy on Metrological Traceability of Measurement Results

General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.