How to set calibration intervals

There is no universal interval. Set it from risk, manufacturer data and — above all — the instrument's own as-found history.

ValiTrac AI editorialUpdated 2026-09-132 min read

An interval is a decision about how long you are prepared to trust an instrument between checks. Guidance from ILAC and OIML describes methods, from fixed intervals adjusted on results to statistical control charts, but all of them rest on the same evidence: how the instrument behaved last time.

Inputs

  • Manufacturer's stability specification.
  • Required tolerance versus the instrument's typical drift.
  • Severity of use, environment and handling.
  • Consequence of an out-of-tolerance result being discovered late.
  • As-found data from previous calibrations: the strongest input.

Adjusting

A run of in-tolerance as-found results with small drift supports lengthening the interval; an out-of-tolerance result shortens it and triggers an impact assessment. Document the rule you use so that intervals are defensible rather than habitual.

Frequently asked questions

How often should a temperature data logger be calibrated?
Annually is the most common starting point for loggers in GDP use; the interval should then be adjusted from as-found results and the consequence of an undetected drift.
Can a calibration interval be longer than one year?
Yes, if as-found history shows the instrument stays well within tolerance across the interval and the risk assessment supports it. ILAC-G24 describes methods for adjusting intervals on evidence.
Does the manufacturer set the calibration interval?
The manufacturer's stability specification is an input; the interval is the user's decision under ISO/IEC 17025 clause 6.4 and GxP expectations.