How temperature data loggers are calibrated
Loggers are calibrated as a system — sensor, electronics and firmware — usually in a chamber, at the points that matter for their use.
A data logger's reading is the output of its sensor, its analogue electronics and its firmware. Calibrating the sensor alone would miss the rest, so loggers are normally calibrated as a whole, reading their logged values against a reference thermometer placed alongside them.
Typical approach
- Environmental chamber (or bath for probe-type loggers) held at each calibration point until stable.
- Reference thermometer close to the loggers, at the same height and airflow.
- Calibration points spanning the use range — for 2–8 °C use, typically around 2, 5 and 8 °C or wider.
- Sufficient dwell time for the logger's response, then several logged readings compared with the reference.
- Corrections and uncertainty reported per point.
Watch for
Chamber uniformity and stability dominate the uncertainty; the logger's own resolution (often 0.1 °C) and response time matter; and the calibration must be at the points the logger will be used at — a calibration at 25 °C says little about 4 °C.
Frequently asked questions
- How is a wireless temperature logger calibrated?
- The same way as any logger: placed in a stable chamber or bath enclosure with a reference thermometer, allowed to stabilise, and its logged readings compared with the reference at each point.
- What uncertainty is typical for data logger calibration?
- Around 0.1–0.3 °C at k = 2 for 0.1 °C-resolution loggers, dominated by resolution and chamber variation; better for high-resolution loggers in a good bath.
References
- [1]EURAMET cg-20: Calibration of Climatic Chambers
- [2]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories
General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.
Have a question on this topic?
Ask ValiTrac AI and see the evidence and calculation behind the answer.
Ask ValiTrac AIFree calculator: Guard band / decision rule
Turn a tolerance and an expanded uncertainty into acceptance limits under ILAC-G8 rules, and see the false-accept and false-reject risks.
Open the calculatorCalibration on ValiTrac
Support calibration workflows, uncertainty budgets and certificate generation.
See the calibration workflowRelated articles
Dry block vs environmental chamber
Two common ways to generate a temperature for calibration, suited to different sensors, ranges and uncertainties.
How to choose a temperature data logger for mapping and monitoring
Match resolution and accuracy to the acceptance criteria, check the calibration offer, and look at memory, battery, export format and response time before price.