Ice point, triple point of water and other fixed points

Fixed points realise defined temperatures without a calibrated reference. The ice point (0 °C, ~±0.002 °C) is accessible to any laboratory; the triple point of water (0.01 °C) is the anchor of ITS-90.

ValiTrac AI editorialUpdated 2026-09-132 min read

The ice point

  • Shaved or crushed ice from pure water, packed in a Dewar and saturated with cold water so no free water sits at the bottom.
  • Probe immersed to adequate depth, allowed to equilibrate; the reading should be stable to within resolution.
  • Re-pack as the ice melts; a slush with floating ice is not an ice point.
  • Common errors: tap water, large ice cubes, insufficient immersion, probe touching the Dewar.

Triple point of water

A sealed glass cell with pure water, ice and vapour in equilibrium defines 273.16 K exactly in ITS-90. It is used to measure R(0.01 °C) of a reference PRT before and after use, so its drift can be tracked and its coefficients corrected. Preparation requires an ice mantle and a maintenance bath; the reward is a reference check with negligible uncertainty.

Frequently asked questions

How often should a reference PRT be ice-point checked?
Before and after each calibration session is common practice; at minimum, on a schedule that catches drift well within the calibration interval.