Dew point and relative humidity calculator
Convert between temperature, relative humidity and dew point, and see how much a temperature error shifts an RH reading.
The sensitivity table answers the calibration question: if the sensor's temperature reading is wrong by ±0.2, ±0.5 or ±1 °C while the moisture content is unchanged, what RH does it report?
Result
| Temperature error | Reads (−) | Reads (+) | RH error |
|---|---|---|---|
| ±0.2 °C | 50.6 % | 49.4 % | ≈ ±0.6 %rh |
| ±0.5 °C | 51.6 % | 48.5 % | ≈ ±1.6 %rh |
| ±1.0 °C | 53.2 % | 47.0 % | ≈ ±3.2 %rh |
Runs entirely in your browser · nothing you enter is sent to a server
Magnus formula over liquid water with Alduchov–Eskridge constants; about 0.4 % in vapour pressure between −40 and 50 °C. Not for frost point over ice.
A worked example
A calibration laboratory holds 21.0 °C and 45 % relative humidity. Items arrive from a 5 °C refrigerator, and the question is whether they will wet on the bench.
What you start with
- Air temperature
- 21.0 °C
- Relative humidity
- 45 %
The arithmetic
- 01
Get the saturation vapour pressure at the air temperature, then scale it by the humidity.
e = 0.45 × e_s(21.0 °C)
- 02
Solve the same relation backwards for the temperature at which that vapour pressure saturates.
dew point = 8.61 °C
- 03
Take the sensitivity of the humidity reading to air temperature at this point.
∂RH/∂t = −2.76 %RH per °C
Dew point 8.61 °C. Anything brought into the room colder than 8.6 °C will condense.
The item from the 5 °C refrigerator wets, and no amount of wiping fixes a resistance measurement made on a damp sensor. The sensitivity figure is the other half: the humidity reading moves 2.8 points for every degree the air temperature moves, so a specification of ±5 %RH is in practice a specification of ±1.8 °C — which is both easier to control and far easier to measure traceably.
How the calculation works
- 01Saturation vapour pressure over water: e_s(T) = 6.1094 exp(17.625 T / (243.04 + T)) hPa.
- 02Dew point: T_d = b γ / (a − γ), where γ = ln(RH/100) + a T/(b + T), a = 17.625, b = 243.04 °C.
- 03Absolute humidity = 216.7 e / (T + 273.15) g/m³, with e the actual vapour pressure in hPa.
- 04RH sensitivity at constant vapour pressure: d(RH)/dT = −RH a b / (b + T)².
Limitations
- Valid over liquid water from about −40 °C to +50 °C; not for frost point (over ice).
- Magnus is an approximation (about 0.4 % in e_s); precision hygrometry uses Sonntag or Wexler equations with enhancement factors.
Frequently asked questions
- What is the dew point at 20 °C and 50 % RH?
- About 9.3 °C. Cooling the air to that temperature at constant moisture content brings it to saturation.
- How much does a 1 °C temperature error change RH?
- At 20 °C and 50 % RH, roughly 3 %rh per kelvin — a 1 °C error in the sensor's temperature reads as about 3 %rh. This is why humidity calibration starts with thermal equality between sensor and reference.
Need the reasoning, not just the number?
Ask ValiTrac AI and see the standards evidence and the engine calculation behind the answer.
Ask ValiTrac AIRead the theory
Humidity monitoring in storage areas: when and how
Monitor relative humidity where product, packaging or a label requires it. The sensors drift, and their uncertainty is larger than you get for temperature.
Calibrating relative-humidity sensors
Saturation vapour pressure climbs about 6 % per degree, so half a degree between sensor and reference costs more RH error than most whole budgets.
Qualifying a stability or climatic chamber
Chambers are qualified for temperature and humidity uniformity, stability and recovery — with acceptance criteria tied to the ICH conditions they will run.
Other calculators
- Type A uncertainty
- En number
- Protocol builder
- Mapping analyser
- Calibration interval
- Thermal lag simulator
- Mapping placement planner
- TUR calculator
- Calibration curve
- Uncertainty budget
- Guard band / decision rule
- MKT & logger data check
- Pt100 converter
- Sampling interval
- Thermocouple mV ↔ °C
- Equipment selector
- Certificate checker
- Laboratory finder
- Temperature converter
- Pressure converter
- 4–20 mA calculator
- Accuracy spec calculator
- NTC thermistor
- Wet-bulb humidity
- Balance minimum weight
- pH slope & offset
- Pt100 lead wire error
- Pressure gauge calibration
- Calibration worksheet
- Qualification workspace