How to calibrate a Pt100
Calibration by comparison against a reference thermometer, and the factors that decide the uncertainty.
A Pt100 is a platinum resistance thermometer with a nominal resistance of 100 Ω at 0 °C. Industrial Pt100 sensors are made to tolerance classes defined in IEC 60751, but a tolerance class is not a calibration: calibration establishes the actual relationship between the sensor's resistance (or the indicated temperature of the instrument it is connected to) and temperature, with an uncertainty.
Calibration by comparison
- Choose calibration points across the range of use, including the extremes.
- Generate each temperature in a suitable medium — a stirred liquid bath, a dry-block calibrator or a chamber — and allow full stabilisation.
- Immerse the Pt100 and the reference thermometer to sufficient depth, close together.
- Record repeated readings of both; the reference must be traceably calibrated.
- Compute the correction (reference minus indicated) at each point and its uncertainty.
Where the uncertainty comes from
Typical contributions are the reference thermometer's calibration uncertainty and drift, the medium's stability and uniformity, immersion effects, self-heating, the readout resolution and, for the unit under test, repeatability and hysteresis. Each is expressed as a standard uncertainty, combined in quadrature, and expanded with a coverage factor (usually k = 2) following the GUM.
Reporting
The certificate should give the measured values, the corrections or deviations, the uncertainty and its coverage factor, the reference standards used with their traceability, and the environmental conditions. If a conformity statement to a tolerance class is made, the decision rule must be stated.
ValiTrac's AI assistant can retrieve evidence on this procedure from its approved sources and, where a numeric combination is needed, calls the deterministic uncertainty engine rather than estimating the arithmetic.
Choosing the points
Points are chosen from use, not from convenience. Cover the working range including both extremes, because extrapolating beyond a calibrated point is not supported by anything on the certificate. Add a point near any limit the instrument has to police: a logger guarding an 8 °C ceiling should be calibrated near 8 °C, not at 0 °C and 50 °C with the interesting region interpolated.
Three to five points is typical for industrial work. A single point establishes an offset and says nothing about the slope; two points cannot reveal curvature. If the readout applies ITS-90 or Callendar–Van Dusen coefficients, enough points are needed to fit them — which is a different and larger requirement than a few correction values.
Lead resistance, and why three wires are not four
A Pt100 changes by about 0.385 Ω per °C near 0 °C, so lead resistance matters enormously. A two-wire connection adds the whole loop resistance to the reading: 1 Ω of cable is 2.6 °C of error, and it is always in the same direction. Three-wire compensation subtracts one current leg from the other and works only to the extent that the two legs are identical; a 0.1 Ω imbalance from a repaired junction box or differing lead lengths leaves about 0.26 °C behind. Four-wire separates current and sense paths entirely and removes the term. For calibration, use four wires, and if the instrument will be used on three in service, calibrate it that way as well and record which.
| Component | Half-width or SD | Divisor | u_i (°C) |
|---|---|---|---|
| Reference certificate, U at k = 2 | 0.020 | 2 | 0.0100 |
| Reference drift over the interval | 0.010 | √3 | 0.0058 |
| Bath gradient between reference and UUT | 0.020 | √3 | 0.0115 |
| Bath stability during readings | 0.008 | 1 | 0.0080 |
| Self-heating, residual after correction | 0.005 | √3 | 0.0029 |
| UUT resolution, 0.1 °C display | 0.050 | √3 | 0.0289 |
| UUT repeatability, n = 10 | 0.012 | 1 | 0.0120 |
| Combined u_c | 0.0353 | ||
| Expanded U at k = 2 | 0.07 |
Resolution dominates, which is the usual finding for a 0.1 °C logger and the reason a more expensive reference thermometer would barely move this number.
Mistakes that cost a calibration
- Insufficient immersion. The largest avoidable error in the whole procedure, and invisible unless you test for it by immersing 20–30 mm further and watching the reading.
- Reading before stabilisation. Take readings only when both the reference and the unit under test have stopped trending, and record several rather than one.
- Ignoring self-heating. Check it by halving the measurement current; the difference is three quarters of the full effect.
- Entering certificate coefficients into the readout and then treating the readout as contributing nothing. It has its own accuracy, resolution and drift.
- Confusing a tolerance class with a calibration. Class A is a manufacturing limit on a new sensor; it says nothing about the one in your hand after two years in a cold room.
Frequently asked questions
- What equipment is needed to calibrate a Pt100?
- A stable temperature source (stirred liquid bath, dry block or chamber), a traceably calibrated reference thermometer with its readout, and a calibrated readout for the Pt100 under test.
- How many points should a Pt100 calibration have?
- At least three across the range of use, including both ends and any temperature where a decision is made; near 0 °C is useful for PRTs because the ice point provides an independent check.
- What uncertainty can be achieved when calibrating a Pt100?
- From a few millikelvin with an SPRT reference in a good bath to 0.05–0.1 °C in a dry block with an industrial reference; the readout resolution of the unit under test often sets the floor.
References
- [1]IEC 60751 — Industrial platinum resistance thermometers and platinum temperature sensors
- [2]JCGM 100:2008 — Evaluation of measurement data: Guide to the expression of uncertainty in measurement (GUM)
- [3]EURAMET cg-13: Calibration of Temperature Block Calibrators
General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.
Have a question on this topic?
Ask ValiTrac AI and see the evidence and calculation behind the answer.
Ask ValiTrac AIFree calculator: Certificate checker
Paste the text of a calibration certificate and screen it against ISO/IEC 17025:2017 clause 7.8: uncertainty with k, traceability, as-found and as-left, the decision rule behind a pass, and every other required element, with the matched text shown for each.
Open the calculatorCalibration on ValiTrac
Support calibration workflows, uncertainty budgets and certificate generation.
Open the calibration worksheetRelated articles
Dry block vs environmental chamber
Two common ways to generate a temperature for calibration, suited to different sensors, ranges and uncertainties.
Measurement uncertainty explained
What an uncertainty statement means, how it is built up, and why a coverage factor matters.