Pt100 tolerance classes explained (IEC 60751)

Classes AA, A, B and C widen with temperature: Class A is ±0.15 °C at 0 °C but ±0.55 °C at 200 °C. A tolerance class is not a measurement uncertainty.

EquipmentValiTrac AI editorialUpdated 2026-10-083 min read

IEC 60751 defines a single resistance-versus-temperature curve that every platinum resistance thermometer is meant to follow, then sets out how far an individual sensor is permitted to deviate from it. The tolerance classes are those permitted deviations. They are a property of the manufactured part, not of a measurement you make with it.

The four classes

Each is the fixed term plus a fraction of the absolute temperature, so the band is tightest at the ice point and widens in both directions.

Tolerance limits in °C, computed from the IEC 60751 expressions. A dash marks temperatures outside the class's defined range.
t (°C)AAABCNominal R
−200——±1.30±2.6018.52 Ω
−100—±0.35±0.80±1.6060.26 Ω
−50±0.18±0.25±0.55±1.1080.31 Ω
0±0.10±0.15±0.30±0.60100.00 Ω
100±0.27±0.35±0.80±1.60138.51 Ω
200±0.44±0.55±1.30±2.60175.86 Ω
300—±0.75±1.80±3.60212.05 Ω
450—±1.05±2.55±5.10264.18 Ω
600——±3.30±6.60313.71 Ω
Tolerance limits in °C, computed from the IEC 60751 expressions. A dash marks temperatures outside the class's defined range.

The expressions are ±(0.10 + 0.0017|t|) for AA, ±(0.15 + 0.002|t|) for A, ±(0.30 + 0.005|t|) for B and ±(0.60 + 0.01|t|) for C. Class C is exactly twice class B at every temperature, because both terms double. Class B is not twice class A: they share a doubled fixed term but the proportional terms are 0.005 against 0.002, so B pulls further ahead as the temperature rises, from twice A at 0 °C to about 2.4 times A at 450 °C.

Ranges matter as much as the numbers

  • Class AA: −50 to +250 °C for wire-wound elements, extended to +600 °C for thin film.
  • Class A: −100 to +450 °C.
  • Class B and C: −196 to +600 °C.

This catches people out in cold-chain work. A ULT freezer running at −80 °C is outside the defined range of both class AA and class A, so a supplier quoting "class A" for a −80 °C application is quoting a figure the standard does not define there. Class B is the only class defined that low, at ±0.70 °C for −80 °C.

What the tolerance is worth in ohms

Converting through the sensitivity makes the size of these bands concrete. At 0 °C a Pt100 changes 0.3908 Ω/°C, so class A's ±0.15 °C is ±0.059 Ω. At 300 °C the sensitivity has fallen to 0.3562 Ω/°C and class A's ±0.75 °C is ±0.267 Ω. If your readout resolves 0.01 Ω, it resolves roughly a fortieth of the class A band at the ice point: the sensor, not the instrument, is the limit.

Tolerance is not uncertainty

This is the point the classes are most often used to dodge. A class A sensor is guaranteed to lie somewhere inside ±0.15 °C of the standard curve at 0 °C. It does not tell you where inside, it says nothing about the instrument reading it, the lead resistance, self-heating or the immersion, and it says nothing about drift after a year in service.

If you treat the class as an uncertainty you are making a rectangular assumption over the band, which gives u = 0.15/√3 = 0.087 °C at 0 °C for class A, before any other component. That is defensible as a bound for an uncalibrated sensor, and it is usually much worse than the same sensor's uncertainty once calibrated, which is the whole argument for calibrating.

A certificate supersedes the class entirely. Once a sensor has been calibrated, use the certificate's correction and uncertainty. Quoting both a class and a certificate for the same sensor, and then choosing whichever is more convenient, is the wrong way round.

Frequently asked questions

What is the tolerance of a class A Pt100?
±(0.15 + 0.002|t|) °C, over −100 to +450 °C. That is ±0.15 °C at 0 °C, ±0.35 °C at 100 °C and ±1.05 °C at 450 °C. The band widens with temperature in both directions from the ice point.
What is the difference between class A and class B?
Class B is ±(0.30 + 0.005|t|) against class A's ±(0.15 + 0.002|t|). They are in a 2:1 ratio only at 0 °C, ±0.30 against ±0.15 °C; because B's proportional term is 2.5 times A's, the gap widens with temperature, reaching ±1.30 against ±0.55 °C at 200 °C. Class B is also defined over a wider range, down to −196 °C.
Can I use a class A Pt100 at −80 °C?
You can use the sensor, but the class does not apply there. Class A is defined from −100 °C, and class AA only from −50 °C. For ULT work the relevant class is B, and in practice the sensor should be calibrated at its working temperature rather than relied on by class.
Is class AA always better than class A?
It is tighter where it is defined, ±0.10 °C against ±0.15 °C at 0 °C, but it is defined only from −50 to +250 °C for wire-wound elements. Outside that range class AA is not an option, so it is not a general upgrade.
Can I use the tolerance class as my measurement uncertainty?
Only as a bound for an uncalibrated sensor, treated as rectangular: class A at 0 °C gives u = 0.15/√3 = 0.087 °C. It covers the sensor alone and nothing else in the chain. Once the sensor is calibrated, the certificate replaces the class.

References

  1. [1]IEC 60751:2022 — Industrial platinum resistance thermometers and platinum temperature sensors
  2. [2]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories

General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.