Pt100 tolerance classes explained (IEC 60751)
Class AA, A, B and C define the maximum permissible deviation of a platinum resistance thermometer from the standard curve as a function of temperature. They are tolerances, not uncertainties.
IEC 60751 specifies the resistance–temperature relationship for industrial platinum resistance thermometers (R₀ = 100 Ω, α = 0.003851 /°C) and the tolerance classes. The tolerance grows with |t| because the deviation from the nominal curve accumulates away from 0 °C.
Worked values
- At 0 °C: AA ±0.10, A ±0.15, B ±0.30 °C.
- At 100 °C: AA ±0.27, A ±0.35, B ±0.80 °C.
- At −40 °C: AA ±0.17, A ±0.23, B ±0.50 °C.
Using the class in a budget
For an uncalibrated sensor, the class tolerance is a rectangular Type B limit. For a calibrated sensor the class is irrelevant: the certificate's correction and uncertainty apply. A Class B sensor calibrated at the points of use can be a perfectly good 0.05 °C instrument — until it drifts.
Frequently asked questions
- Is a Pt1000 covered by the same classes?
- Yes; IEC 60751 covers platinum sensors with other nominal resistances using the same tolerance classes.
References
- [1]IEC 60751:2022 — Industrial platinum resistance thermometers and platinum temperature sensors
General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.
Have a question on this topic?
Ask ValiTrac AI and see the evidence and calculation behind the answer.
Ask ValiTrac AIFree calculator: Pt100 converter
Convert between resistance and temperature for Pt100 / Pt1000 sensors, with class AA, A, B and C tolerances at that temperature.
Open the calculatorCalibration on ValiTrac
Support calibration workflows, uncertainty budgets and certificate generation.
See the calibration workflowRelated articles
How to calibrate a Pt100
Calibration by comparison against a reference thermometer, and the factors that decide the uncertainty.
Self-heating in resistance thermometers
The measuring current warms the sensing element. The effect depends on the medium — so calibrate at a current and condition representative of use.