Maintaining calibration equipment between calibrations
References drift, baths contaminate, inserts wear, batteries die. Maintenance and intermediate checks find those before the next calibration does.
ISO/IEC 17025 clause 6.4 requires procedures for handling, transport, storage, use and planned maintenance of equipment, and intermediate checks where needed to maintain confidence in performance. For temperature equipment the plan is short but must be followed.
A minimal plan
- Reference PRT: ice-point check before each use or weekly; record and chart the value; investigate a shift of more than the certificate uncertainty.
- Dry block: annual re-characterisation of gradients; insert inspection for wear and contamination.
- Liquid bath: fluid level, contamination and viscosity checked; fluid change interval defined.
- Readouts: zero and reference-resistor checks.
- Loggers: battery replacement schedule; clock check against a reference time.
- All: transport cases, temperature limits for storage, and a rule that a dropped reference is checked before reuse.
A control chart of ice-point readings is the single most informative maintenance record a temperature laboratory keeps.
A maintenance plan you can actually run
Clause 6.4 requires procedures for handling, transport, storage, use and planned maintenance, and intermediate checks where they are needed to maintain confidence. The failure mode is a plan so elaborate nobody follows it. This one fits on a page.
| Equipment | Check | Frequency | Trigger to act |
|---|---|---|---|
| Reference PRT | Ice point, recorded and charted | Monthly, or before critical work | Shift larger than the certificate uncertainty at 0 °C |
| Readout or bridge | Stable reference resistor of known value | Monthly | Change beyond its specification |
| Dry block | Compare against the reference at one set point | Quarterly | Difference beyond the characterised gradient |
| Dry block | Full re-characterisation of gradients, loaded | Annually, and after any insert change | Any term worse than the budget assumes |
| Liquid bath | Uniformity and stability at working points | Annually, and after a fluid change | As above |
| Bath fluid | Visual and condition check | Monthly | Darkening, thickening, water uptake, growth |
| Working thermometers | Compare against the reference at one point | Set from their own history | Approaching tolerance |
The ice-point chart earns its place
Of everything in that table, the control chart of ice-point readings is the single most informative record a temperature laboratory keeps. It costs a few minutes a month and it does three jobs: it detects a drifting or damaged reference long before the annual certificate does, it provides the evidence for lengthening or shortening the calibration interval, and — on the day an out-of-tolerance finding triggers a look-back — it locates when the change began, turning a twelve-month impact assessment into a one-month one.
The unglamorous failures
- Batteries. A logger or handheld that dies mid-measurement takes the data with it, and rechargeable cells fail gradually rather than suddenly. Replace on a schedule, not on failure.
- Well inserts. They wear, corrode and pick up debris, and a worn insert opens the air gap that dominates a dry block's radial term. Inspect, and re-characterise after replacement.
- Connectors and leads. A corroded terminal produces a lead-resistance imbalance that a three-wire connection converts straight into a temperature offset.
- Transport. A reference PRT carried in a van is the commonest cause of a sudden ice-point shift. Use a proper case, and check the ice point after any journey.
- Bath fluid ageing. Ethanol takes up water and its freezing point rises; oil oxidises and thickens; water grows things. All three change the uniformity you characterised.
Record the checks even when they find nothing. A file containing only the month a problem appeared reads as though the checks began that month, and it cannot support an interval decision.
Frequently asked questions
- What if an intermediate check shows a shift?
- Quarantine the reference, evaluate the impact on results since the last good check under the nonconforming-work procedure, and recalibrate.
References
- [1]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories
General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.
Have a question on this topic?
Ask ValiTrac AI and see the evidence and calculation behind the answer.
Ask ValiTrac AIFree calculator: Pt100 converter
Convert between resistance and temperature for Pt100 / Pt1000 sensors, with class AA, A, B and C tolerances at that temperature.
Open the calculatorCalibration on ValiTrac
Support calibration workflows, uncertainty budgets and certificate generation.
Open the calibration worksheetRelated articles
The ice-point check
A properly made ice bath gives 0 °C to within a few millikelvin — a cheap, repeatable way to watch a PRT for drift between calibrations.
Equipment requirements in ISO/IEC 17025 (clause 6.4)
Equipment must be fit for purpose, calibrated where it affects results, checked, protected from adjustment, and fully recorded — including software.