Choosing monitoring locations from mapping results
Permanent sensors go where the mapping found the worst case — the hot spot, the cold spot, the slowest-recovering position — not where the cable reaches.
The purpose of a mapping study is to find where a space behaves worst. Routine monitoring is then placed at those positions so that if the worst location is within limits, everything else is too. A monitoring sensor placed at the control probe's position typically shows the best-controlled point in the room — the least informative place to look.
Selection rules
- Hot spot (highest mean or maximum) and cold spot (lowest mean or minimum) from the loaded mapping.
- Positions with the slowest recovery after door openings or defrost.
- Each distinct zone if the space has more than one HVAC regime.
- Product-height positions, not ceiling or floor, unless product is stored there.
Record the rationale
A monitoring plan should show, for each sensor, which mapping position it corresponds to and why. When the space changes and is re-mapped, the plan is reviewed against the new results.
From mapping result to monitoring plan
The selection is mechanical once the study is done, which is the point of doing the study. Each monitoring sensor should trace to a mapping position and a reason.
| Sensor | Mapping position | Why it was chosen |
|---|---|---|
| M1 | S07, top shelf near the door | Hot spot by maximum; the upper-limit risk |
| M2 | S02, low shelf in the evaporator discharge | Cold spot; the freezing risk, which nothing else would see |
| M3 | S11, geometric centre | Slowest recovery after door openings; represents the bulk of the volume |
| M4 | Outside the room | Ambient reference, which makes every event above interpretable |
Two sensors would be a defensible minimum here. Four is better and the marginal cost is small. What would not be defensible is one sensor at the centre, because the centre is the best-behaved point in the room and a monitoring record from it will say the room is fine while the door shelf runs at 10 °C.
Both extremes, not just the warm one
Monitoring plans routinely cover the hot spot and ignore the cold spot, because upper-limit excursions are the ones people think about. In a 2–8 °C room the freezing risk is the more serious of the two: it is irreversible for many biologicals, it happens in the evaporator discharge where nobody puts a sensor, and an averaged room reading will never show it. If the mapping found a position below the lower limit, that position needs either a monitoring sensor or a physically enforced exclusion zone.
Recording the rationale, and keeping it current
- The monitoring plan names, for each sensor, the mapping position it corresponds to and the reason. Without that link the two datasets are unrelated and neither supports the other.
- Where an existing sensor is not at a worst-case position, the mapping report has to say so and recommend the move — and somebody has to make it. A recommendation never implemented documents that you knew.
- When the space is re-mapped, review the plan against the new results. Racking changes move the worst case, and the sensor stays where it was bolted.
- Cover each distinct zone. A large space with separate airflow regions needs a sensor per region, not one sensor and an average.
- Keep one ambient reference. It costs a channel and explains most of what the others do.
Frequently asked questions
- Can the mapping loggers stay in place as the monitoring system?
- They can, if they are calibrated for that role, connected to alarms, and their data are captured in a compliant system. Usually dedicated monitoring sensors are installed at the chosen positions.
- How many monitoring sensors does a cold room need?
- At least two (hot and cold spot) for a small unit; larger rooms add sensors per zone and near doors. The mapping results and risk assessment set the number.
References
- [1]EU Guidelines on Good Distribution Practice of medicinal products for human use (2013/C 343/01)
- [2]WHO Technical Supplement 8 to TRS 961 Annex 9: Temperature mapping of storage areas (2015)
- [3]WHO Technical Supplement 16 to TRS 961 Annex 9: Environmental temperature monitoring systems for fixed storage areas
General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.
Have a question on this topic?
Ask ValiTrac AI and see the evidence and calculation behind the answer.
Ask ValiTrac AIFree calculator: MKT & logger data check
Paste logger readings from storage or a cargo shipment: MKT per ICH Q1A(R2), time out of range, and which figure actually governs the excursion.
Open the calculatorRelated articles
Sensor placement principles for mapping studies
Place sensors where product is, where risk is, and where the monitoring probe might go — and write down why.
Alarm limits vs acceptance limits: setting them correctly
Acceptance limits come from the product; alarm limits sit inside them to give time to act, alert limits inside those. Uncertainty and delay set the gaps.