Choosing calibration points

Calibrate where you measure. Points should bracket the use range and include any temperature where a decision is made.

ValiTrac AI editorialUpdated 2026-09-132 min read
  • Cover the range of use, including both ends; do not extrapolate beyond the calibrated points.
  • Include the specific temperatures at which the instrument makes decisions — alarm limits, storage set points.
  • Use enough points to reveal non-linearity; three is a minimum for a simple sensor, more for wide ranges.
  • For PRTs, include or approach 0 °C where a known reference point exists.
  • Match the medium and immersion to service conditions where practical.

A logger used only in a 2–8 °C refrigerator does not benefit from a certificate at 25 °C and 50 °C; it needs points at and around its limits.

Frequently asked questions

What calibration points should a 2–8 °C logger have?
Typically around 2, 5 and 8 °C, or slightly wider (0, 5, 10 °C) so the use range is bracketed and the alarm limits are covered.
What calibration points for a −20 °C freezer logger?
Points that bracket the storage condition, for example −25, −20 and −15 °C, plus any alarm limit temperature.