Reporting uncertainty: rounding and significant figures

Report the expanded uncertainty to at most two significant figures, round the result to the same decimal place, and state the coverage factor.

MetrologyValiTrac AI editorialUpdated 2026-10-093 min read

An expanded uncertainty of 0.11734 °C claims knowledge no budget can support. The GUM recommends reporting U to at most two significant figures, and the measured value rounded to the decimal place of U. Rounding of intermediate values is avoided; rounding happens once, at the end.

Examples

  • Correction +0.0473 °C, U = 0.0316 °C → report +0.047 °C, U = 0.032 °C (k = 2).
  • Reading 21.8367 °C, U = 0.12 °C → report 21.84 °C ± 0.12 °C (k = 2).

Form of words

The GUM suggests a form such as: 'The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k = 2, providing a level of confidence of approximately 95 %.' Accredited certificates carry an equivalent statement.

The rule, and why it exists

GUM §7.2.6 asks for the expanded uncertainty to be given to at most two significant figures, with the reported value rounded to the same decimal place. The reason is not tidiness. An uncertainty is itself an estimate, built from other estimates, several of which came from a specification sheet or an engineering judgement. Quoting it to five figures claims a precision about the uncertainty that no budget in this field can support, and it is one of the things an assessor will notice first because it signals that the numbers were copied out of a spreadsheet rather than read.

Computed against reported
As computedAs reportedNote
+0.0473 °C, U = 0.03162 °C+0.047 °C, U = 0.032 °C (k = 2)Two figures on U; value matched to it
21.8367 °C, U = 0.1204 °C21.84 °C ± 0.12 °C (k = 2)U to two figures fixes the value at two decimals
−0.0081 °C, U = 0.0702 °C−0.008 °C, U = 0.070 °C (k = 2)The trailing zero is significant; keep it
1.7 °C, U = 0.96 °C1.7 °C, U = 1.0 °C (k = 2)Rounding up across the decade is normal, not an error
Computed against reported

Round once, at the end

Every intermediate value carries full precision until the final report. Rounding each component before combining introduces an error that compounds through the quadrature sum, and rounding u_c before expanding it is the commonest version. Carry the digits, expand, then round the pair together.

The exception worth knowing is GUM §7.2.6's own allowance: where rounding down would reduce the uncertainty by more than about 5 %, round up instead. Reporting 0.0996 as 0.099 is fine; reporting a computed 1.04 as 1.0 when the difference matters is the case the allowance covers.

The statement that goes with the number

A number without a coverage factor cannot be compared with any other number. The GUM's recommended form, which accredited certificates carry in some variation, is: the reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k = 2, providing a level of confidence of approximately 95 %.

  • Always state k, or the coverage probability, or both. A bare ± figure is ambiguous by a factor of two or more.
  • Say approximately 95 %, not 95 %. The coverage is approximate unless effective degrees of freedom were evaluated and k chosen accordingly.
  • Give the unit of the uncertainty explicitly, and say whether it is absolute or relative. Relative uncertainties quoted as a bare percentage next to a temperature are a recurring source of confusion.
  • Keep the result and its uncertainty at the same decimal place throughout the document, including in tables and summaries, so a reader is never asked to reconcile two versions of the same figure.

Frequently asked questions

Should U be rounded up or to nearest?
Round to nearest unless doing so reduces U by more than a few percent, in which case rounding up is the conservative convention.

References

  1. [1]JCGM 100:2008 — Guide to the expression of uncertainty in measurement (GUM)
  2. [2]ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories

General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.