Building an uncertainty budget step by step

Write the model, list the inputs, evaluate each as a standard uncertainty, apply sensitivity coefficients, combine in quadrature, expand with k. One table.

UncertaintyValiTrac AI editorialUpdated 2026-10-113 min read

The eight steps

  • 1. State the measurand precisely: e.g. the correction to the indicated temperature of the unit under test at 5 °C.
  • 2. Write the model: correction = t_ref + δt_ref_drift + δt_bath_gradient + δt_bath_stability − t_uut − δt_uut_resolution … each δ a correction with an expected value of zero and an uncertainty.
  • 3. List the inputs and gather information: certificate uncertainties, specifications, repeatability data, resolutions.
  • 4. Evaluate each as a standard uncertainty (Type A from data; Type B from limits with a distribution).
  • 5. Determine sensitivity coefficients from the model.
  • 6. Combine: u_c² = Σ (c_i u_i)².
  • 7. Check effective degrees of freedom if any Type A component is based on few observations.
  • 8. Expand: U = k u_c, usually k = 2, and report with the value rounded consistently.

A minimal temperature example

Reference certificate 0.02 °C (k=2 → 0.010); reference drift ±0.01 (rect → 0.0058); bath gradient ±0.02 (rect → 0.0115); bath stability observed SD 0.008; UUT resolution 0.1 (rect → 0.0289); UUT repeatability 0.012. Combined ≈ 0.035 °C; U ≈ 0.07 °C at k = 2. The resolution term dominates — a common finding for 0.1 °C loggers.

The same example as a budget table: Pt100 logger at 5 °C, comparison in a stirred bath
ComponentValueDistributionDivisoru_i (°C)Contribution to u_c²
Reference certificate, U at k = 20.020Normal20.01008 %
Reference drift since calibration±0.010Rectangular√30.00583 %
Bath spatial gradient±0.020Rectangular√30.011511 %
Bath temporal stability (observed SD)0.008Normal10.00805 %
UUT display resolution 0.1 °C±0.050Rectangular√30.028968 %
UUT repeatability (Type A, n = 10)0.012Normal10.012012 %
Combined u_c0.0351100 %
Expanded U, k = 20.0702
The same example as a budget table: Pt100 logger at 5 °C, comparison in a stirred bath

Read the table, not just the answer

The last column is the point of the exercise. Resolution contributes 68 % of the variance — more than everything else combined — because a 0.1 °C display is a ± 0.05 °C rectangular half-interval, and squaring makes large terms dominate. Buying a better reference thermometer here would move the total from 0.0351 to about 0.0339: a 3 % improvement for real money. Reading the logger at a finer resolution, or interpolating across the display step, is worth ten times as much.

That is the general lesson. A budget is a diagnostic, not a certificate decoration. Sort by contribution and work on the top row.

Rounding, and the mistake almost everyone makes

The combined uncertainty above is 0.0351 °C and the expanded is 0.0702 °C. Report it as U = 0.07 °C, not 0.0702 °C. GUM §7.2.6 asks for the uncertainty to be given to at most two significant figures, and the result rounded to the same decimal place. Quoting four figures of uncertainty claims a precision about the uncertainty itself that the estimate cannot support — and an assessor will notice.

The correction is then reported alongside it at the matching resolution: for instance correction +0.03 °C, U = 0.07 °C (k = 2, approximately 95 %). Round at the end, once, never at each intermediate step.

This article is the method. The uncertainty budget calculator is the arithmetic: enter each component with its distribution and sensitivity coefficient, and it applies the divisors, combines in quadrature, computes effective degrees of freedom and reports U — including which component dominates the variance, which is usually the only output that changes what you do next.

Checks before you sign it

  • Does every input have a sensitivity coefficient, even where it is 1? Writing c_i = 1 deliberately is different from forgetting it exists.
  • Is anything counted twice? Reference drift and reference certificate overlap if the certificate already covers the interval.
  • Is anything missing that you know about? Immersion, self-heating and hysteresis are the usual omissions in temperature work.
  • Are the divisors right? Rectangular is √3, triangular √6, U-shaped √2, and a certificate at k = 2 divides by 2 — not by √3.
  • Does the answer pass a sanity check against the tolerance? A U comparable to the tolerance means the conformity statement needs a decision rule.

Frequently asked questions

How many components should a budget have?
As many as the model needs and no more. Six to twelve is typical for a temperature calibration; a budget with fifty components usually double-counts.
Should I include components that are negligible?
List them with their evaluated values so the reviewer can see they were considered; they will contribute nothing to the combined value.

References

  1. [1]JCGM 100:2008 — Guide to the expression of uncertainty in measurement (GUM)
  2. [2]EA-4/02 M: Evaluation of the Uncertainty of Measurement in Calibration

General technical guidance written against the cited sources. It is not regulatory or legal advice and does not replace the applicable standard, guideline or a qualified reviewer's judgement.