Temperature calibration in practice

Bench-level technique for PRTs, thermocouples and loggers: sources, immersion, self-heating, hysteresis, references and the ice point.

11 articles · about 37 min · for Laboratory technicians and engineers performing or specifying temperature calibrations.

  1. 01CalibrationHow to calibrate a Pt100Calibration by comparison against a reference thermometer, and the factors that decide the uncertainty.4 min
  2. 02CalibrationChoosing calibration pointsCalibrate where you measure. Points should bracket the use range and include any temperature where a decision is made.3 min
  3. 03CalibrationLiquid bath versus dry-block calibratorWith the same reference and the same repeatability, a stirred bath gives U = 0.038 °C and a dry block 0.139 °C, where the source becomes the largest term.3 min
  4. 04CalibrationImmersion depth and stem conductionA probe conducts heat along its stem. Insufficient immersion makes the sensor read closer to room temperature than to the medium.3 min
  5. 05CalibrationSelf-heating in resistance thermometersThe measuring current warms the element above its surroundings: under a millikelvin in a stirred bath, tens in still air. Calibrate in the medium you use.4 min
  6. 06CalibrationHysteresis in temperature sensorsA sensor can read differently at the same temperature depending on where it came from. Calibrations that only approach from one direction hide it.3 min
  7. 07CalibrationSelecting a reference thermometerPick a reference for stability first and certificate uncertainty second: one that drifts between its own calibrations is worse than a larger honest number.4 min
  8. 08CalibrationThe ice-point checkA properly made ice bath gives 0 °C to within a few millikelvin — a cheap, repeatable way to watch a PRT for drift between calibrations.3 min
  9. 09UncertaintyThe reference thermometer's contribution to uncertaintyThe reference contributes its certificate uncertainty, drift since calibration, and any interpolation or self-heating effect. Together they set the floor.3 min
  10. 10CalibrationData logger calibration: how temperature loggers are calibratedA logger is calibrated as a whole system, sensor and electronics and firmware together, in a chamber, at the temperatures you actually use it at.3 min
  11. 11CalibrationHow thermocouples are calibratedComparison against a reference in a bath or furnace, with the reference junction realised. Ignore it on a type K at 420 °C and you read 398 °C.4 min

Questions while you read?

Ask ValiTrac AI — answers cite the standards and show any calculation.

Open ValiTrac AI